Section 7: Command reference Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-356 2651A-901-01 Rev. A / March 2011
Details
Releases a trigger that was asserted with an indefinite pulse width, as well as a trigger that was latched in
response to receiving a synchronous mode trigger.
Example
tsplink.trigger[3].release()
Releases trigger Line 3.
Also see
tsplink.trigger[N].assert() (on page 7-350)
tsplink.trigger[N].clear()
(on page 7-351)
tsplink.trigger[N].mode
(on page 7-352)
tsplink.trigger[N].overrun
(on page 7-354)
tsplink.trigger[N].pulsewidth
(on page 7-355)
tsplink.trigger[N].stimulus
(on page 7-357)
tsplink.trigger[N].wait()
(on page 7-358)
tsplink.trigger[N].reset()
This function resets some of the TSP-Link
®
trigger settings to their factory defaults.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
tsplink.trigger[N].reset()
N
The trigger line (1 to 3)
Details
The tsplink.trigger[N].reset() function resets the following attributes to their factory defaults:
• tsplink.trigger[N].mode
• tsplink.trigger[N].stimulus
• tsplink.trigger[N].pulsewidth
This also clears tsplink.trigger[N].overrun.
Also see
tsplink.trigger[N].pulsewidth (on page 7-355)
tsplink.trigger[N].mode
(on page 7-352)
tsplink.trigger[N].overrun
(on page 7-354)
tsplink.trigger[N].stimulus
(on page 7-357)