Model 2651A High Power System SourceMeter® Instrument Reference Manual Section 7: Command reference
2651A-901-01 Rev. A / March 2011 7-139
Usage
serialno = localnode.serialno
serialno
The serial number of the instrument
Details
This read-only attribute indicates the instrument serial number.
Example
display.clear()
display.settext(localnode.serialno)
Clears the unit's display.
Places the unit's serial number on the top line of its display.
Also see
localnode.description (on page 7-133)
localnode.model
(on page 7-134)
localnode.revision (on page 7-138)
localnode.showerrors
This attribute sets whether or not the instrument automatically sends generated errors.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Power cycle Not saved 0 (disabled)
Usage
errorMode = localnode.showerrors
localnode.showerrors = errorMode
errorMode
Enables (1) or disables (0) show-errors state
Details
If this attribute is set to 1, the instrument automatically sends any generated errors stored in the error queue, and
then clears the queue. Errors are processed after executing a command message (just before issuing a prompt,
if prompts are enabled).
If this attribute is set to 0, errors are left in the error queue and must be explicitly read or cleared.
When using this command from a remote node, localnode should be replaced with the node reference, for
example node[5].showerrors.
Example
localnode.showerrors = 1
Enables sending of generated errors.
Also see
localnode.prompts (on page 7-136)