Details
Set this attribute to zero (0) to disable the automatic trigger output.
Do not use the stimulus attribute for generating output triggers under script control. Use
digio.trigger[N].assert() instead.
Trigger stimulus for a digital I/O line may be set to one of the existing trigger event IDs, described in the following
table.
Trigger event IDs*
Event ID Event description
smuX.trigger.SWEEPING_EVENT_ID
Occurs when the SMU transitions from the idle state
to the arm layer of the trigger model
smuX.trigger.ARMED_EVENT_ID
Occurs when the SMU moves from the arm layer to
the trigger layer of the trigger model
smuX.trigger.SOURCE_COMPLETE_EVENT_ID
Occurs when the SMU completes a source action
smuX.trigger.MEASURE_COMPLETE_EVENT_ID
Occurs when the SMU completes a measure action
smuX.trigger.PULSE_COMPLETE_EVENT_ID
Occurs when the SMU completes a pulse
smuX.trigger.SWEEP_COMPLETE_EVENT_ID
Occurs when the SMU completes a sweep
smuX.trigger.IDLE_EVENT_ID
Occurs when the SMU returns to the idle state
digio.trigger[N].EVENT_ID
Occurs when an edge is detected on a digital I/O line
tsplink.trigger[N].EVENT_ID
Occurs when an edge is detected on a TSP-Link line
lan.trigger[N].EVENT_ID
Occurs when the appropriate LXI trigger packet is
received on LAN trigger object N
display.trigger.EVENT_ID
Occurs when the TRIG key on the front panel is
pressed.
trigger.EVENT_ID
Occurs when a *TRG command is received on the
remote interface.
(GPIB only) Occurs when a GET bus command is
received.
(VXI-11 only) Occurs with the VXI-11 command
device_trigger. Reference the VXI-11 standard
for additional details on the device trigger operation.
trigger.blender[N].EVENT_ID
Occurs after a collection of events is detected.
trigger.timer[N].EVENT_ID
Occurs when a delay expires.
* Use the name of the trigger event ID to set the stimulus value rather than the numeric value. Using the name
makes the code compatible for future upgrades (for example, if the numeric values must change when
enhancements are added to the instrument).
Example 1
digio.trigger[3].stimulus = 0
Clear the trigger
stimulus of digital
I/O line 3.
Example 2
digio.trigger[3].stimulus =
smua.trigger.SOURCE_COMPLETE_EVENT_ID
Set the trigger
stimulus of digital
I/O line 3 to be the
source complete
event.

Table of Contents

Other manuals for Keithley 2651A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2651A and is the answer not in the manual?

Keithley 2651A Specifications

General IconGeneral
BrandKeithley
Model2651A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals