Model 2651A High Power System SourceMeter® Instrument Reference Manual Appendix E: Status model
2651A-901-01 Rev. A / March 2011 E-9
Figure 134: Operation status trigger overrun registers
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status LAN
Summary Register
+
status.operation.instrument.lan
Connection (CON)
Configuring (CONF)
To LAN Bit (LAN) in Operation Status
Instrument Summary Register
(status.operation.instrument)
Trigger Overrun (TRGOVR)
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status LAN
Trigger Overrun Register
+
status.operation.instrument.lan.trigger_overrun
LAN1 (LAN1)
LAN2 (LAN2)
LAN3 (LAN3)
LAN4 (LAN4)
LAN5 (LAN5)
LAN6 (LAN6)
LAN7 (LAN7)
LAN8 (LAN8)
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status Trigger
Overrun Summary
+
status.operation.trigger_overrun
SMU A (SMUA)
LAN (LAN)
To Trigger Overrun Bit (TRGOVR) in
Operation Status Register
(status.operation)
TSP-Link (TSPLINK)
Digital I/O (DIGIO)
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status SMU A
Trigger Overrun Register
+
status.operation.instrument.smua.trigger_overrun
Arm Overrun (ARM)
Source Overrun (SRC)
Measure Overrun (MEAS)
End Pulse Overrun (ENDP)
To Trigger Overrun Bit (TRGOVR) in
Operation Status SMU A Summary Register
(status.operation.instrument smua)
Trigger Timer (TRGTMR)
Trigger Blender (TRGBLND)

Table of Contents

Other manuals for Keithley 2651A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2651A and is the answer not in the manual?

Keithley 2651A Specifications

General IconGeneral
BrandKeithley
Model2651A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals