EasyManua.ls Logo

Keithley 2651A

Keithley 2651A
800 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 2651A High Power System SourceMeter® Instrument Reference Manual Appendix E: Status model
2651A-901-01 Rev. A / March 2011 E-9
Figure 134: Operation status trigger overrun registers
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status LAN
Summary Register
+
status.operation.instrument.lan
Connection (CON)
Configuring (CONF)
To LAN Bit (LAN) in Operation Status
Instrument Summary Register
(status.operation.instrument)
Trigger Overrun (TRGOVR)
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status LAN
Trigger Overrun Register
+
status.operation.instrument.lan.trigger_overrun
LAN1 (LAN1)
LAN2 (LAN2)
LAN3 (LAN3)
LAN4 (LAN4)
LAN5 (LAN5)
LAN6 (LAN6)
LAN7 (LAN7)
LAN8 (LAN8)
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status Trigger
Overrun Summary
+
status.operation.trigger_overrun
SMU A (SMUA)
LAN (LAN)
To Trigger Overrun Bit (TRGOVR) in
Operation Status Register
(status.operation)
TSP-Link (TSPLINK)
Digital I/O (DIGIO)
0
6
5
3
4
2
1
7
14
13
12
11
10
9
8
15
Operation Status SMU A
Trigger Overrun Register
+
status.operation.instrument.smua.trigger_overrun
Arm Overrun (ARM)
Source Overrun (SRC)
Measure Overrun (MEAS)
End Pulse Overrun (ENDP)
To Trigger Overrun Bit (TRGOVR) in
Operation Status SMU A Summary Register
(status.operation.instrument smua)
Trigger Timer (TRGTMR)
Trigger Blender (TRGBLND)

Table of Contents

Other manuals for Keithley 2651A

Related product manuals