Millipore System Documentation Set
186 Operating and Maintenance Manual Publication XITXSP121 P Rev.A
10. The calibration coefficients 1, 2, 3 and the regression coefficient for each transducer
will be updated automatically if adjustments are necessary.
11. At the completion of the external calibration, an “internal calibration computation” is
automatically performed to match the secondary pressure transducer and pressure
regulator to the primary transducer.
12. The results of the calibration are displayed on the screen ( See Figure 75, pg184 )
13. When the test is complete print the report by clicking
.
14. If the test must be stopped before completion, click
to terminate the internal
calibration in progress.