Millipore System Documentation Set
100 Operating and Maintenance Manual Publication XITXSP121 P Rev.A
Table 9: Tab F6, HydroCorr Test Entries
Entry Optional or
Mandatory
Instructions Notes
Minimum Gas
Volume
Mandatory Enter a value from
50 ml to 2000 ml
It is important to control the
initial gas volume during a
HydroCorr test. The
Minimum and Maximum gas
volume inputs insure that the
test is run within acceptable
gas volume limits.
Maximum Gas
Volume
Mandatory Enter a value from
100 ml to 4000 ml
HydroCorr Test Time Mandatory Enter a value from
5–60 minutes
Unlike the diffusion test, the
HydroCorr test does run for
the predetermined test time
that is entered.
HydroCorr Test
Pressure
Mandatory Enter a value from
5–90 psig
This is the pressure at which
the HydroCorr test is
conducted. It is specified by
the filter manufacturer.
HydroCorr Test
Specification
Mandatory Enter a value from
0.03–10 ml/min
This is the HydroCorr flow
rate specification for the filter
catalogue number being
entered. It is specified by the
filter manufacturer.
HydroCorr Constants Mandatory Enter the appropriate
values for each
constant.