EasyManua.ls Logo

Millipore Integritest Exacta - Pneumatic Connections

Millipore Integritest Exacta
284 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Millipore System Documentation Set
28 Operating and Maintenance Manual Publication XITXSP121 P Rev.A
Pneumatic Connections
Diagram Key: A. Gas Inlet (see Gas Supply Requirements below)
B. Exhaust
C. From Housing
D. To Housing
Figure 4: Side Panel
Gas Inlet (quick disconnect labeled “Air Inlet”)
Exhaust (tubing connector)
Outlet (dual quick disconnects labeled “From Housing” and “To Housing”)
Gas Supply Requirements
The instrument must be connected to a regulated supply of clean dry air (>-20 C dew point) or
nitrogen at 35–120 psig (2.4–8.5 bar).
Note: To minimize the likelihood of water contamination, it is recommended that the inlet
gas be filtered with an external coalescing filter.
Gas Inlet Connection
To Gas Supply
Do not remove the gas inlet filter or damage to the instrument may result.
Connect a regulated supply of clean dry air or nitrogen, 35–120 psig (2.4–8.5
bar), to the inlet tubing assembly of the instrument using the ¼” NPTF side of the
gas filter.
Supply at least 15 psi (1.02 bar) more than the maximum pressure required to
perform the filter test.
To Instrument
Slide quick disconnect fitting of Integritest Exacta inlet tubing assembly onto the
connector stem labeled Air Inlet.
Grasp the rear of the connector and push it firmly against the connector stem. Do
not grasp the flared section when making a connection. This section must be able
to slide freely into place
Release it when you hear it click into place.

Table of Contents

Related product manuals