Integritest Exacta Filter Integrity Test Instrument
Operating and Maintenance Manual References 265
REFERENCES
Recommended HydroCorr Gas Volumes
Table 34: Recommended HydroCorr Gas Volumes
Filter Size Initial System Gas
Volume Range (cm
3
)
Preferred Initial System
Gas Volume (cm
3
)
50 mm capsule (such as Aervent-50) 20–60 50
Millidisk cartridge (10–40 stack) 30–100 50
4-inch capsule (such as Opticap) 30–100 50
4-inch cartridge (such as Optiseal) 30–100 50
5-inch cartridge 30–100 50
pleated cartridge filters (in
10-inch multiples)
one 10-inch element 100–600 300
two 10-inch elements (one 20-inch
cartridge or two 10-inch cartridges)
200–700 450
three 10-inch elements (one 30-inch
cartridge or three 10-inch cartridges, etc.)
300–1000 600
four 10-inch elements 350–1100 750
five 10-inch elements 400–1200 900
six 10-inch elements 450–1300 1050
seven 10-inch elements 500–1400 1200
eight 10-inch elements 550–1500 1350
nine 10-inch elements 600–1500 1500
Conversion Factors
Table 35: Conversion Factors
Convert Value (Metric) Value (Imperial)
Pressure 1 kPa (Kilopascal) 0.145 psi (Lbs/Sq.Inch)
6.894 kPa (.06894 bar) 1 psi
Temperature Degrees Celsius Degrees Fahrenheit
Conversion Formulas ˚C x 9/5 +32 = ˚F (˚F–32 ) x 5/9 =˚C
Length 1 Meter 3.28 Feet (39.36 in.)
.3048 Meter 1 Foot