Millipore System Documentation Set
220 Operating and Maintenance Manual Publication XITXSP121 P Rev.A
Table 27: Status Error Messages
Status
Error
No.
Status Window
Message
Short Description Detailed Description
1 MATH ERROR General
Failure
General error in operation
Used to convey errors occurring during operation
that include, improper set up parameters,
computational errors, etc.
2 INPUT ERROR Failed
Precondition
Check
Error that occurs at the initiation of a test. The
input parameters and the test & math dll‘s are
checked for validity prior to initiating the test.
This error condition will arise if a condition fails.
3 OUTPUT
ERROR
Failed
Postcondition
Check
Error that occurs at the completion of a test. The
output termination parameters and are checked
for validity at the completion of the test. This
error condition will arise if a condition fails.
4 INPUT TOO
SMALL
Failed Param
Low Range
Check
Failed minimum boundary value for the input
value listed.
5 INPUT TOO
HIGH
Failed Param
High Range
Check
Failed maximum boundary value for the input
value listed.
6 NETWORK
ERROR
RPC Failure
In network communications the RPC (Remote
Protocol Control) layer is the set of functions that
allow an IT Exacta hardware control software to
communicate with a remote User Interface (either
in an IT Exacta or on another computer). This
error message is generated when there is a
failure in this communications link and the two
are not connected.
7 NETWORK
UNAVAILABLE
RPC Service
Not Available
The Remote Communications Control package is
being used by another computer and unavailable
at this time.
8 GENERAL
ERROR
INSUFFI-
CIENT GAS
PRESSURE
Test Execution
General
Failure
General fault condition associated with an error
in the test protocol. Generally caused by incorrect
starting values being transmitted to the hardware
for either test operation or system configuration.
See the operations log for details.
Insufficient inlet gas pressure to run a test.