Figure 76: Temperature Calibration
3. The following dialog box will appear.
4. Enter the reference gauge identification number. This will be included on the
calibration printout.
5. Place the internal temperature sensor and the referee temperature sensor in the first
bath, and allow about 3 minutes for equilibration.
6. Hit the
button.