Integritest Exacta Filter Integrity Test Instrument
Operating and Maintenance Manual Maintenance and Calibration 193
7. The instrument will take 7 temperature readings from the internal temperature sensor
and then prompt the operator to enter the value from the referee sensor. Upon input,
the instrument will take the 8
th
reading
8. Repeat this process for each of the three calibration temperatures.
9. After the third calibration, a “Bath #4” dialogue screen will appear. Hit the
button immediately.
10. The calibration constants 1, 2, 3 and the regression constant for temperature sensor(s)
will be updated automatically if adjustments are necessary, and displayed on the
screen.
11. When the test is complete, print the report by clicking on the Print Icon
.