Millipore System Documentation Set
222 Operating and Maintenance Manual Publication XITXSP121 P Rev.A
Status
Error
No.
Status Window
Message
Short Description Detailed Description
18 CALIBRATE Out Of
Calibration
This error indicates that the transducer or
pressure regulator is out of calibration when
compared to the manufacturer specifications. A
mechanical adjustment or replacement is
required.
19 USER
TERMINATED
User Aborted
Test
User termination.
20 NO FILTER Filter Not
Found
The filter being requested is not found in the filter
database. Both the filter catalogue number and
revision number needs to match that being
requested.
21 NO HC
INPUTS
HC Inputs Not
Found
Specific to a HydroCorr test, the HydroCorr input
values could not be found.
22 SETUP
INSTRUMENT
Setup Inputs
Not Found
Setup inputs such as test pressure, minimum
bubblepoint, test time could not be found in the
database.
23 NO ERROR
LOG
Client Log
Error
There has been an error in accessing the log file
associated with the user interface.
24 UNDEFINED
USER
User Not
Found
Log in error – the user name is not valid for this
network
General Messages
Messages may appear during any instrument operation. Table 28 lists these messages with
their cause and location.