Internal Assemblies — Front Chassis 4: Repair
129
6 If present, disconnect the NBP airtube from the NBP Port.
7 If present, remove the SpO
2
PCA, but do not move the Rear Chassis. Follow the instructions for
“SpO2 PCA” on page 126 starting from the “Removal” section.
8 Open the ECG Port connector and disconnect the ECG Port from the Processor PCA, see Figure 84
(Rear Chassis not shown for clarity).
9 Disconnect the CO
2
cable from the Processor PCA, see Figure 84.
10 Remove the CO
2
cable core stabilizer screw and the stabilizer, see Figure 84.
11 Remove the Measurement Module fork screw, see Figure 85.
12 Gently pull the Measurement Module fork away from the Rear Chassis to free the rib from the lock
and lift up, see Figure 85 on page 130.
13 Gently sliding the CO
2
cable under the Processor PCA, remove the Measurement Module assembly
from the Front Case.
Figure 84 Measurement Module CO
2
Connections
CO
2
cable, PCA side
CO
2
cable, port side
CO
2
cable stabilizer
ECG connector
NBP Port