Plug-in Modules 5 Repair and Disassembly
113
NOTE After replacing the Pump, reset the displayed value displayed using the Reset PumpOpTime selection
(Service Mode>CO
2
Setup). When the PumpOpTime has been reset an INOP will be generated:
CO
2
OCCLUSION. To clear this INOP you must perform a flow check and store the flow in
Service Mode (select Store Flow)
Refit Procedures for the Measurement Server Extension
Tools Required:
A thin-bladed screwdriver.
A pair of large tweezers.
In addition, for refitting the Pump, you will need a large-bladed screwdriver.
WARNING There is high voltage inside the Instrument (800V). Do not connect the Measurement Server
Extension to a Monitor while the Extension housing is open.
As well, parts inside the instrument may be contaminated with bacteria. protect yourself from possible
infection by wearing examination gloves during these procedures.
Refitting the CO
2
Scrubber
WARNING The CO
2
Scrubber contains lithium hydroxide monohydrate. This is a strong base. Do not open or
damage the CO
2
Scrubber. If you come into contact with the CO
2
Scrubber material, flush the area
immediately with water and consult a doctor.
To refit the CO
2
Scrubber, do the following:
1 O
2
Scrubber through the bracket to meet the Extension intake tube.
2 Push the intake tube firmly into the scrubber end to connect it.
3 Holding the body of the CO
2
Scrubber with tweezers, feed the CO
2
Scrubber fresh air intake
under the second bracket and position it.
Refitting the Pump
To refit the Pump, do the following:
1 Gently connect the power lead to the Extension.
NOTE The power lead can only be connected one way.Do not try to force the power lead into position.
Instead, align it correctly and connect it gently.
2 Connect the flow tubing to the Pump.
NOTE Be sure to reconnect the inlet tube to the inlet valve and the outlet tube to the outlet valve.
1 Being careful not to touch anything else in the Extension, insert the pump into the bracket on the
PC board. Make sure that the pump is horizontal and does not touch the PC board. (Vibration
from the pump in operation will damage the Extension if the pump touches the PC board.)
2 Replace the top part of the bracket and screw firmly into position.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP70 and is the answer not in the manual?

Philips IntelliVue MP70 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP70
CategoryMedical Equipment
LanguageEnglish

Related product manuals