Troubleshooting Guide 4 Troubleshooting
75
Individual Parameter INOPS
If any of the following parameter INOP messages are issued try the respective parameter in another
device. If the INOP message persists replace the parameter module, the MMS or other indicated
device.
•CO
2
EQUIP MALF
• ECG EQUIP MALF
• NBP EQUIP MALF
• P1 EQUIP MALF
• RESP EQUIP MALF
•SpO
2
EQUIP MALF
•SpO
2
TRANSDUC MALF
•SvO
2
EQUIP MALF
•tcpO
2
(or tcpCO
2
) EQUIP MALF
• T1 EQUIP MALF
• VueLnk EQU. MALFI