Troubleshooting Guide 4 Troubleshooting
75
Individual Parameter INOPS
If any of the following parameter INOP messages are issued try the respective parameter in another
device. If the INOP message persists replace the parameter module, the MMS or other indicated
device.
•CO
2
EQUIP MALF
ECG EQUIP MALF
NBP EQUIP MALF
P1 EQUIP MALF
RESP EQUIP MALF
•SpO
2
EQUIP MALF
•SpO
2
TRANSDUC MALF
•SvO
2
EQUIP MALF
•tcpO
2
(or tcpCO
2
) EQUIP MALF
T1 EQUIP MALF
VueLnk EQU. MALFI