Troubleshooting the Anesthetic Gas Module 9 Anesthetic Gas Module
225
Jumpers
The following table shows the correct jumper settings for the O
2
sensor (if installed).
The top cover Jumper, JP7, must always be closed, whether or not the top cover is connected.
-XPSHU 3DUDPDJQHWLF2
6HQVRU 1R2
6HQVRU
10IRQ1
OPEN OPEN
10IRQ2
OPEN OPEN
10IRQ3
CLOSED OPEN