Tests 3 Testing and Maintenance
47
Tools required: Ohmmeter.
1 Plug an M8087-61001 cable into the Nurse Call Relay connector.
2 Connect the ohmmeter and measure the pins as indicated in the diagram and table.
3 The relay contacts should behave as described above. The behavior may vary depending on
configuration choices. See the Configuration Guide for details on Alarm Relay settings.
ECG Sync Performance Test
This test checks the performance of ECG synchronization between the monitor and a defibrillator. It
only needs to be performed when this feature is in use as a protocol at the customer site.
Tools required:
• Defibrillator with ECG Sync and Marker Output.
• Patient simulator.
1 Connect the patient simulator to the ECG connector on the Measurement server and the
defibrillator to the ECG Sync Output on the monitoring.
2 Set the patient simulator to the following configuration:
– HR = 100 bpm.
– ECG sinus rhythm.
3 Switch the defibrillator to simulation mode.
4 Check that the marker pulse is displayed before the T-wave begins.
VueLink Tests using VueLink Test Module
Use the VueLink plug-in test module ( M1186-60510) to test M1032A VueLink modules.
Test Procedure
You must preselect the test module to ON in Configuration Mode. Therefore, the test module must be
one of the devices made available for selection during configuration of the VueLink module.
Carry out the test itself in monitoring mode. For information concerning the configuration of
VueLink modules see the M1032A VueLink Module Handbook.
1 Plug the VueLink module into the FMS.
2 Press the Setup key on the front of the VueLink module.
3 In the Setup VueLink menu select Device, then select Test Module.
4 Select Confirm to store the selection and wait for the message “Switched to new device”.
5 Plug in the test module.
6 Connect the modules by plugging one end of the cable (part number M1032-61661) into the
connector on the front of the VueLink Module, and the other end into the connector on the front
of the Test Module.
7 Select the wave segment on the screen, where you want the waves to appear. In the wave menu,
select
Change Wave, then select WAVE.
8 Select the VueLink SmartKey, then select the TEST Plug-In pop-up key