EasyManua.ls Logo

Philips IntelliVue MP70 - Connectors

Philips IntelliVue MP70
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9 Anesthetic Gas Module Troubleshooting the Anesthetic Gas Module
224
Connectors
The following table lists the connectors on the main PCB.
TP10
M1026A A/D Converter Input
TP11
IR source voltage for agent ID head
(R180)
7.92 V ±20mV
TP13
Isolated Ground for isolated RS-232
communications with the Host Computer
via J14
TP14
+12VHP +12V ±600mV
TP15
+5V +5V ±250mV
TP16
-12V -12V ±600mV
TP17
+12V +12V ±600mV
TP18
Analog Ground 3 (AGND3)
1. Measurement requires an oscilloscope.
7HVW3RLQW 'HVFULSWLRQ 7ROHUDQFH9DOXH
Connector Description
J1 IR Measurement Assembly Head
J2 Agent ID Head
J3 Factory use only
J4 Factory use only
J5 DC Power: pins 1&2: +12VHP, pins 3&4: +5V GND (DGND), pin 6:
AGND, pin 7: AGND, pin 8: -12V, pin 9: +12V
J6 PC Board connector (also for NVRAM data transfer)
JP7 Top cover jumper
J8 Pump driver for DC pump and AC pump (DC PUMP/AC PUMP)
1
1. Use DC pump only.
J10 Oxygen Transducer Connection
J13 Factory use only
J14 Connection to Host Computer for communications, RS-232
J15 Fan power (FAN1)
J16 Not used
J17 Power LED
J19 Factory use only
J20 Factory use only
J21 O
2
Solenoid (SOL2)
J22 Optional Front Panel I/O

Table of Contents

Related product manuals