9 Anesthetic Gas Module Repairing the Anesthetic Gas Module
230
The Top Cover
Removal
To remove the top cover (refer to Figure 54 and Figure 55):
1 Make sure that the module is switched off and isolated from the mains power supply.
2 Remove the watertrap from the front of the cover (to avoid fluids in the watertrap reaching the
water fuses when the cover is tipped).
3 Using a cross-tipped screwdriver, remove the 7 screws (1) securing the top cover to the body.
These screws are located at the rear of the module and on the sides.
4 Slide the top cover (2) forward approximately 4cm.
NOTE At this stage, the top cover is still connected to the main PC board by a flat cable and the internal
Nafion tubing.
5 Carefully lift the top cover until the flat cable connector (3) leading to the main PC board is
accessible.
6 Remove the flat cable connector (3) from the top cover.
7 If necessary, remove the internal Nafion tubing.
8 Remove the top cover from the module.
Replacement
To replace the top cover (refer to Figure 56 and Figure 57):
1 Ensure that the module is switched off and isolated from the mains power supply.
2 Reconnect the flat cable connector (3).
3 If necessary, reconnect the internal Nafion tubing, following the given markings.
4 Carefully lower the top cover (4) onto the chassis (5).
5 Slide the top cover towards the rear of the module until the locating holes on the top cover are
aligned with the threaded bores in the module.
6 Using a cross-tipped screwdriver, replace the 7 screws (1) securing the top cover to the module.
The locating holes are under the top cover at the rear and sides of the module.