Remote Device Site Requirements 8 Site Preparation
159
Multi-Measurement Server M3001A or Flexible Module Server
M8048A
Space Requirements Multi-Measurement Server M3001A
Size (W x D x H)
188.0mm x 96.5 mm x 51.5 mm
(7.40 x 3.80 x 2.03)
Weight
650g (1.4 lb)
Space Requirements Flexible Module Server M8048A
Size (W x D x H)
320 mm x 120mm x 35mm (12.6 x 4.72 x 5.3)
Weight
< 3500g (7.7lb)
Environmental Requirements Multi-Measurement Server M3001A
Temperature
Operating: 0 to 45
º
C (32 to 113
º
F)
Storage: -40 to 70
º
C (-40 to 158
º
F)
Humidity
Operating: 95% relative humidity (RH) max. @ 40
º
C (104
º
F)
Storage: 90% relative humidity (RH) max. @ 65
º
C (150
º
F)
Altitude
Operating: -500m to 4600m (-1600 to 15000 ft.)
Storage: -500m to 15300m (-1600 to 50000 ft.)
Environmental Requirements Flexible Module Server M8048A
Temperature
Operating: 0 to 45
º
C (32 to 113
º
F)
Storage: -40 to 70
º
C (-40 to 158
º
F)
Humidity
Operating: 95% relative humidity (RH) max. @ 40
º
C (104
º
F)
Storage: 90% relative humidity (RH) max. @ 65
º
C (150
º
F)
Altitude
Operating: -500m to 4600m (-1600 to 15000 ft.)
Storage: -500m to 15300m (-1600 to 50000 ft.)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP70 and is the answer not in the manual?

Philips IntelliVue MP70 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP70
CategoryMedical Equipment
LanguageEnglish

Related product manuals