8 Site Preparation Monitor M8005A and M8007A Site Requirements
156
Providing conduit and/or trunking of a sufficient cross-sectional area for the planned cables and
possible future expansion (for additional components or systems). See Cabling Options and Conduit
Size Requirements for cable specifications for remote installations.
Providing and/or installing suitable wall boxes to accommodate the faceplates.
Pulling Cables
WARNING NEVER run power cables through the same conduit or trunking used for system cables.
Installing Wall Boxes
It is the customer's responsibility to provide and install wallboxes to house faceplates. The customer
must notify the Philips installation coordinator of which size is to be used.
Procedures for Philips Personnel
Before you begin the procedures in the installation sections, ensure that the customer has completed all
necessary preparations outlined in the previous section, "Procedures for Local Staff."
Monitor M8005A and M8007A Site Requirements
Space Requirements
The situating of the monitor should be planned such that the nursing staff are able to monitor the
patient with relative ease, with all patient connectors and controls readily available and the displays
clearly visible. The location should also allow access to service personnel without excessive disruption
and should have sufficient clearance all round to allow air circulation.
Maximum dimensions and weight:
Size (W x H x D)
405mm x 360mm x 170mm (15.95 x 14.17 x 6.69)
without handle and Speedpoint device
Weight
10kg (22.05lb) without options
Environmental Requirements
The environment where the MP60/MP70 monitor will be used should be reasonably free from
vibration, dust and corrosive or explosive gases. The ambient operating and storage conditions for the
MP60/MP70 monitor must be observed. If these conditions are not met, the accuracy of the system
will be affected and damage can occur.
Temperature
Operating: 0 to 35
º
C (32 to 95
º
F)
Storage: -20 to 60
º
C (-4 to 140
º
F)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP70 and is the answer not in the manual?

Philips IntelliVue MP70 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP70
CategoryMedical Equipment
LanguageEnglish

Related product manuals