EasyManuals Logo

Philips IntelliVue MP70 Service Manual

Philips IntelliVue MP70
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
Tests 3 Testing and Maintenance
39
Leakage check
Pump check
Flow check and calibration, if required.
Noise check
CO
2
Cal check and calibration, if required.
CO
2
Cal verification using 2nd cal gas
Perform all checks in the same session.
Barometric Pressure Check and Calibration
Check the barometric pressure value in the sidestream CO
2
extension as follows:
1 Go into service mode and select Setup CO
2
menu.
2 Connect a FilterLine to the sidestream CO
2
input. This activates the pump in the sidestream CO
2
Extension.
3 The status line at the bottom of the screen displays CO2 pressure reading (ambient/cell) xxx/yyy
where xxx is the ambient pressure and yyy is the measured cell pressure. Check whether the
ambient pressure value (x1) matches (within the acceptable tolerance of ±12mm Hg) the reference
value you have received. If so, proceed to the leakage check. If the value is not correct, calibrate as
follows.
a. Select
CO
2
then select Barom.Press to activate a table of values.
b. Select the value in the table which matches the reference value received from a reliable local
source (airport, regional weather station or hospital weather station). (The values are displayed
with a resolution of 2 mmHg up to 500 mmHg and a resolution of 1 mmHg from 500 mmHg
to 825 mmHg.) Note: the selected value must be within ±10% of the current measured ambient
pressure, otherwise an error message will occur at restarting the monitor.
c. Confirm the barometric pressure setting.
d. Check that the ambient pressure displayed in the status line at the bottom of the screen is the
same as the value which you selected from the list in step b.
Leakage Check
The leakage check consists of checking the tubing between:
the pump outlet and the measurement server extension outlet and
the pump inlet and FilterLine inlet.
Check the users guide of the flowmeter for details on how to make a correct flow reading.
Part 1
1 Go into service mode and select Setup CO
2
menu.
2 Connect a FilterLine to the sidestream CO
2
input to start the pump running.
3 Check the ambient pressure and the cell pressure shown in the monitors status line. The cell
pressure should be approximately 20 mmHg lower than ambient pressure.
4 Connect the flowmeter outlet to the FilterLine inlet using a flexible connecting tube.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP70 and is the answer not in the manual?

Philips IntelliVue MP70 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP70
CategoryMedical Equipment
LanguageEnglish

Related product manuals