9 Anesthetic Gas Module Checking and Calibrating the Anesthetic Gas Module
186
Checking and Calibrating the Anesthetic Gas
Module
This chapter explains how to check the Anesthetic Gas Module to ensure that it is operating within its
specified limits. A list of the equipment required to carry out the checks is included, as well as step-by
step instructions for the calibrations.
If you receive fail indications while testing, refer to the troubleshooting section of this chapter for
guidance. If you are instructed to remove or replace parts of the Anesthetic Gas Module refer to the
respective section.
Access Service Functions of the M1026A Anesthetic Gas Module
Enter service mode and select the service screen (see Testing and Maintenance for instructions on
entering service mode). In the
Setup Gas Analyzer menu you can choose whether the Gas
Analyzer Diagnostic
window or the Gas Analyzer Calibration window should be
displayed. In this window you can as well start the flow calibration, the barometric pressure calibration
and the gas span calibration.
The
Setup Gas Analyzer menu can be accessed by either going to the Main Setup menu and
selecting
Gas Analyzer, or by pressing the setup key on the Anesthetic Gas Module.
Figure 38 Gas Analyzer Diagnostic Window
This window provides you with diagnostic information about the AGM. In the
Setup Gas
Analyzer
menu select Service Window then select Calibration to access this window.