Tests Recommended When Performing... 3 Testing and Maintenance
35
Tests Recommended When Performing...
Installation
Repair
Preventive Maintenance
Perform preventive maintenance tests:
• NBP calibration
• Sidestream CO
2
calibration
• Pump and scrubber replacement.
Performance Verifications
Perform all safety, accuracy and performance test procedures listed in the following sections. If a
particular measurement is in question, perform the measurement performance test only.
Upgrades
Service Event
(When performing...
Tes t Bl o c k s Requi red
...Complete these tests)
Installation of monitor with no display connected
to the VGA output
Perform Visual and Power On Test Blocks
Installation of monitor with a display connected
to the VGA output
Perform Visual, Power On and Safety (1) Test
Blocks
Service Event
(When performing...
Tes t Bl o c k s Requi red
...Complete these tests)
Repairs of M3015A Perform Power On and M3015A tests
Repairs where the monitor has been damaged by
impact
Perform Power On and Safety (2) and (3) Test
Blocks
Repairs where the power supply is replaced Perform Safety (2) Test Block
All other IntelliVue Monitoring System repairs Perform Power On Test Block
Service Event
(When performing...
Tes t Bl o c k s Requi red
...Complete these tests)
Hardware and software upgrades Perform Power On Test Block unless otherwise
specified in the Upgrade Installation Notes shipped
with the upgrade.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP70 and is the answer not in the manual?

Philips IntelliVue MP70 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP70
CategoryMedical Equipment
LanguageEnglish

Related product manuals