8 Site Preparation Introduction
154
• Prepare mounts.
• Pull cables, install conduit, install wallboxes.
• Terminate network cables if a Philips Clinical Network is in use.
• It may be necessary to certify the network cable plant, see Philips Clinical Network Installation
Manual for details.
Philips Personnel
• Provide the customer with the safety, environmental and power requirements.
• Assemble mounts.
• Prepare monitor remote cabling.
Procedures for Local Staff
The following tasks must be completed before the procedures for Philips personnel may be started.
Providing Power Outlets
One power outlet for each display and for any peripheral device (for example, a printer or slave display)
is required by the system. Provide a power outlet in the vicinity (1 m or 3 ft) of each component that
requires power.
WARNING Only the power cables provided with the system may be used. For reasons of safety, the use of power
(mains) extension cables or adapters is NOT recommended.
Preparing Mounts
Where ceiling, wall, or shelf mounts are required for mounting the equipment, the customer is
responsible for the following:
• Providing and installing all hardware which is required to install the mounting hardware supplied by
Philips as detailed in the installation notes.
• Making sure that all ceilings, walls, and mounting rails that supports mounting hardware are
suitable for their proposed load.
WARNING It is the customer's responsibility to have the attachment of the mounting hardware to the ceiling, wall,
or mounting rail and the construction of the ceiling, wall, or mounting rail evaluated for structural
integrity and compliance with all local, state and any other required codes by a registered, professional,
structural and/or mechanical engineer.
Although considerable effort has been made to ensure the safety of the ceiling mount installation and
or mounting guidelines, it is to be understood that the installation itself is beyond the control of Philips
Medical Systems. Accordingly, Philips Medical Systems will not be responsible for the failure of any
such installation.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP70 and is the answer not in the manual?

Philips IntelliVue MP70 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP70
CategoryMedical Equipment
LanguageEnglish

Related product manuals