Remote Device Site Requirements 8 Site Preparation
161
Remote Displays (M8031A)
Space Requirements
Size (W x D x H)
With mounting bracket: 333mm x 408mm x 85mm (13.1” x 16” x 3.4”)
With desk stand: 387mm x 408mm x 175mm (15.2” x 16” x 6.9”)
Weight
With mounting bracket: 4900g (10.8lb)
With desk stand: 6900g (15.2lb)
Environmental Requirements
Temperature
Operating: 5 to 45
º
C (41 to 113
º
F
Storage: -20 to 60
º
C (-4 to 140
º
F)
Humidity
Operating: 95% RH max @ 40
º
C (104
º
F)
Storage: 85% RH max @ 50
º
C (122
º
F)
Altitude
Operating: Up to 4600m (15000 ft.)
Storage: Up to 4600m (15000 ft.)
Electrical and Safety Requirements
Voltage ranges:
90V to 264V (13.5A fuse)
Voltage selection:
Wide range input, no voltage selection required
Power consumption:
Maximum ?
Cabling Options and Conduit Size Requirements
The following table describes the cabling options for the M8031A 15” TFT Medical Grade Touch
Display.