EasyManuals Logo

Philips IntelliVue MP70 Service Manual

Philips IntelliVue MP70
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #256 background imageLoading...
Page #256 background image
9 Anesthetic Gas Module Repairing the Anesthetic Gas Module
256
Solenoid Valve #2
Removal
To remove solenoid valve #2 (refer to Figure 78):
1 Ensure that the Module is switched off and isolated from the mains power supply. Remove the top
cover of the module.
2 Remove the pneumatic tubing (1) from solenoid valve #2.
3 Remove the connection (2) from the main PC board.
4 Using a cross-tipped screwdriver, remove the two screws (3) securing the solenoid valve to its
bracket.
5 Remove solenoid valve #2.
Replacement
To replace solenoid valve #2 (refer to Figure 78):
1 Ensure that the Module is switched off and isolated from the mains power supply.
2 Position the solenoid valve on its bracket so that the locating holes on the valve align with threaded
bores in the bracket.
3 Using a cross-tipped screwdriver, replace the two screws (3) securing the solenoid valve to its
bracket.
4 Replace the connection (2) to the main PC board.
5 Replace the pneumatic tubing (1) to solenoid valve #2. Check that all tubing is tightly connected
and show no signs of damage.
NOTE After replacing solenoid valve #2, check that all flat-cable connectors are firmly seated and show no
signs of damage.
6 Replace the top cover of the module.
Now perform the performance checks described Test and Inspection Matrix on page 262.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP70 and is the answer not in the manual?

Philips IntelliVue MP70 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP70
CategoryMedical Equipment
LanguageEnglish

Related product manuals