Troubleshooting Guide 4 Troubleshooting
57
Flexible Module Server Boot Phase
For these steps it is assumed that the Flexible Module Server is connected via SRL-cable to the monitor
Time
(sec.) after
Monitor
Power On
Event
0 Red Error LED switches on immediately
1 Green “Ready” LED switches on
3 Red Error LED is switched off
5 Module Power is switched on
5-8 Module Status LEDs blink once or twice (Module dependent)