EasyManuals Logo

Quantum ODYSSEY HF Series User Manual

Quantum ODYSSEY HF Series
493 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #488 background imageLoading...
Page #488 background image
Appendix D Calibration (Using Generator Service Tools)
Revision W HF Series X-ray Generators - Service Manual
D-61 Quantum Medical Imaging, LLC
and re-enter the observed density value. Continue until the mea-
sured density value equals the Target Gray Level setting to within
acceptable limits. When it does, press Confirm button.
c. The system advances to Calibration Wizard - 60kVp 10cm Calibration
screen.
7. Repeat the previous steps for the 60, 80, 100, 110, 120, and, if 150 kV
system, 130 and 140 kVp calibration points, placing the specified thick-
ness of acrylic in the x-ray beam as indicated on the Calibration Wizard -
AEC kVp Calibration screen for each kV calibration point, until all the
observed density values are within required range.
8. This completes AEC calibration for this particular Receptor/Film Screen.
Repeat AEC calibration procedures for all additional Receptor/Film
Screens that will be used with the system.
9. Verify AEC calibration by taking several exposures in AEC mode and
checking the resulting density value. If necessary, repeat AEC calibration
procedure to further adjust AEC accuracy.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals