Chapter 2 Assembly & Installation
Revision W HF Series X-ray Generators - Service Manual
2-52 Quantum Medical Imaging
•DHCP = Off
IP Address = 10.0.0.71
Subnet mask = 255.255.255.1
Default gateway = 10.0.0.1
Preferred DNS server = 10.0.0.3
10. If necessary, each of these fields can be modified by selecting the field, and
then entering the correct information using the data input Touch Pad that
appears. Note that the touch pad display includes numerical input buttons 0
through 9, including a decimal point, a plus sign (+), a minus sign (-), a clear
(Clr) button, a delete (Del), a Cancel button and an Enter button; however,
erroneous input via the numbers buttons will cause the displayed address to
turn red, indicating that an error has been made either in acceptable value or
entry format. Should an input error occur, it can be corrected by touching the
Del button, followed by touching the correct key. The Clr button clears the
Address window, and allows immediate re-entry of Address values.
11. Touch the Apply button to save new settings.
12. Return to the Generator Help Services Screen by touching the Back to Ser-
vice button at lower left corner of screen.
13. On the Generator Help Services screen, select Generator Service. The
Generator Service Options Menu appears:
Figure 2-25e. General Service Options Menu
ONLY PRESS THESE
BUTTONS WHEN
DIRECTED TO DO SO BY
TECHNICAL SUPPORT

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Quantum ODYSSEY HF Series and is the answer not in the manual?

Quantum ODYSSEY HF Series Specifications

General IconGeneral
BrandQuantum
ModelODYSSEY HF Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals