ABL77 Service Manual Chapter 7: Re-Assembly
Valve Boards – General Information, Continued
Y-cable
installation
(continued)
Step Action
5.
Connect the waste pump cable to the Y-cable connector with the blue
and green wires. Match the polarity of the cables by connecting the
blue and green wires of the Y-cable with the same color wires of the
waste pump cable.
6.
Disconnect the cable connector of valve L3 from J7 of the valve board
in the lower right-hand corner.
7.
Route the L3 valve cable under the L1 valve cable and across the
board, as shown.
8.
Connect the L3 valve cable connector to the Y-cable connector with
the red wires. No polarity match is required with this cable.
Completed
module diagram
Figure 7-7
Continued on next page
Rev. F 7-17