Getting started
R&S
®
TSMA6
32User Manual 4900.8057.02 ─ 11
USB 3.0 (4x, Type A)
See (2) in Figure 3-7.
Connecting external storage devices, data sticks and test mobile phones.
Power limit: max. 900 mA / port
Overall USB current (USB-C, USB 3.0 and USB 2.0): max. 3 A
Docking connector
See (3) in Figure 3-7.
Connector for the battery pack unit R&S TSMA6/B-BP.
DC IN connector
See (4) in Figure 3-7.
Connecting external DC power supply.
Ensure that the voltage and current indicated on the R&S TSMA6 match the available
power source.
Input voltage range: 11 V to 18 V
Input current: max. 8.5 A
Status LEDs
The four status LEDs, [Scanner State], [Scanner Pwr], [Mode], and [Meas] indicated
different states of the R&S TSMA6. For a detailed description, see Chapter 3.2.5, "Sta-
tus LEDs", on page 35.
Power on/off
See (6) in Figure 3-7.
Turns the device on and off.
Coldstart: holding the button > 5 sec.
For details, see Chapter 3.1.8, "Switching on or off R&S TSMA6", on page 25.
Restore button
See (7) in Figure 3-7.
System recovery to factory or user default.
Use a slim, dull object for pressing the button.
Min button hold time for detection: 20 sec
AUX connector (SMA) - synchronization R&S TSME4/6 / R&S TSMExxDC
See (8) in Figure 3-7.
The AUX connector has two functions.