EasyManua.ls Logo

SciCan Statim 2000 - Page 109

SciCan Statim 2000
275 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 107
STAT
IM
2000/2000S Service Guide
96-108639 Rev 3.0
5. STAT
IM
2000S Calibration
STATIM 2000/2000S/2000 G4 Service Guide
10. During sterilization and after all necessary adjustments have been made, the measured
chamber pressure indicated by the reference meter and the STAT
IM
2000S LCD should
not differ by more than 6 kPa.
11. A consecutive failure to achieve the above conditions after adjustment indicates that one
or more of the following components may be faulty:
t Chamber thermocouple
t Pressure sensor
t Pressure interface board
t Controller board (Thermocouple amplifier or A to D converter).
12. When the calibration is complete, press the STOP button on the control box and wait for
the beeps to signal that the cassette may be removed. Press the STOP button again to
reset the unit.
4. Steam generator thermocouple re-calibration
Repeat the steam generator thermocouple calibration procedure in section 2 described above.
5. Switch off the unit and remove all calibration equipment.

Table of Contents

Other manuals for SciCan Statim 2000

Related product manuals