2000/2000S Service Guide
96-108639 Rev 3.0
12. Printer and Data Logger
5. Insert USB flash drive or SD memory card.
6. Power the Data Logger ON.
7. Power the STATIM unit ON.
8. The display on the unit will now show the time and date (not shown when logger not connected)
and the correct time and date will need to be set. To set time, date and unit number, see
Chapter 13, Installation.
9. When this operation is complete, the unit is now ready for use.
Data Logger Troubleshooting
9 pin revision 7.x units only.
If the display reads: “MSD NOT CONNECTED <> INSERT MSD/FLASH”, proceed as follows to
ensure everything is setup correctly.
1. Check the serial cable connection.
2. Check the power connection.
3. Ensure the lower red LED is lit. It is located between the Data Logger’s serial port and power
input.
4. Check that the Mass Storage Device (MSD) is properly inserted in the appropriate slot of the
Data Logger.
5. Repeat the instructions for Installing the Data Logger on your STATIM.
If the display reads: “SAFELY REMOVE MSD <> MSD/FLASH DETECTED”, the MSD/Flash can
be safely removed without affecting the data.
If the display reads: “MSD/FLASH FULL <> REPLACE MSD”, the MSD is full and the data should
be downloaded from the MSD.
If the MSD: has missing lines of Data, refer to the STATIM screen to confirm successful
sterilization.
The Data Logger may be reset by unplugging its power supply, disconnecting the MSD and
waiting 10 seconds. Then re-connect the power adapter and insert the MSD into the Data Logger.
If the problem persists, contact the SciCan Service Center.
STATIM 2000/2000S/2000 G4 Service Guide