2000/2000S Service Guide
96-108639 Rev 3.0
5. STAT
IM
2000S Calibration
STATIM 2000/2000S/2000 G4 Service Guide
5. Attach the pressure reference meter to the pressure tube and the pressure tube to the test
cassette. (Self locking couplings).
6. Install the calibration jumper to controller board pins marked W1.
7. Locate TC ADJ and Vref potentiometers.
1. Control box
2. Controller board
3. SELECT knob
4. VREF potentiometer
5. Digital voltmeter
6. Calibration cassette
7. Thermocouple
8. Digital thermometer
9. TC-ADJ potentiometer
10. Pressure meter
Figure 2