6. Cover Assembly
STATIM 2000/2000S/2000 G4 Service Guide
Reinstalling the Cover
To reinstall the cover, proceed as follows (see Figure 1).
1. Inspect the armature gasket (11) located on the inside front portion of the cover. If it is
damaged or fails to adhere to the surface of the cover, it must be replaced. (See Replacing the
Armature Gasket.)
2. Remove the reservoir cap (5) from the reservoir.
3. Rest the cover (1) on the left side of the unit. Reconnect the keypad and LCD connectors to
controller board.
4. Reconnect the power cord (4) and power the unit ON to check that the LCD and keypad are
functioning. If the amber LED on the front cover is flashing, or the solenoid valve is clicking
repeatedly, the LCD connector is incorrectly installed. Check that the keypad connector
is properly plugged onto the header. Press all of the keypad buttons and check for proper
display messages. After checking, turn the power switch OFF and unplug the power cord.
5. Carefully lift the cover from the work surface. While rotating the cover slide it forward until the
fascia clears the controller board and the front of the armature.
6. Tilt the front of the cover down and the rear of the cover upwards. Carefully reposition the
front portion of the cover back over the front of the controller board (8) and armature. Ensure
that the LEDs at the front left hand corner of the unit fit into the clearance holes in the cover.
BE CAREFUL NOT TO PINCH THE RIBBON CABLES.
7. When the front of the cover is in place, lower the rear portion and gently push it backwards.
Carefully realign the screw holes and reinstall three screws with lockwashers (7) across the
rear of the unit. If the unit being serviced has a bacteria-retentive air filter (9), reinstall the filter.
8. Reinstall the four remaining screws (6), two on each side. Press firmly on the top of the cover
to compress the gaskets and partially reinstall the screws as the holes align. When in place,
tighten all the screws. Do not over tighten the screws.
9. Replace the reservoir filter (if fitted) and fill the reservoir with steam process distilled water then
replace the reservoir cap (5).
10. Reconnect the power cord (4). Select a cycle to ensure that the keypad and LCD function,
then press the STOP key.