Page 257
STAT
IM
2000/2000S Service Guide
96-108639 Rev 3.0
12. Printer and Data Logger
STATIM 2000/2000S/2000 G4 Service Guide
25 pin revision 2.x/5.x/6.x 2000 units only.
Files are stored on the MSD in txt and PDF format. These can be transferred for storage on a
computer hard drive as follows:
1. Ensure the STATIM is in standby. THE UNIT MUST NOT BE RUN WHEN REMOVING THE MSD.
2. Remove MSD from the logger.
3. Insert the MSD in the relevant port on the computer and wait until the computer has recognised
the new device (if transferring for the first time).
4. The contents of the MSD should now be available for viewing in the appropriate application (e.g.
Windows Explorer).
5. Locate, or create the preferred folder for storing the data.
6. Data can now be transferred by either:
a. Drag and Drop
b. Cut and Paste
c. Copy and Paste (if you wish to retain the data on the MSD)
7. Safely ‘eject’ the MSD as appropriate to your computer and remove the MSD.
8. Ensure the STATIM is in standby and re-insert the MSD back in to the Data Logger. THE UNIT
MUST NOT BE RUN WHEN REMOVING THE MSD. Tip: Allow a few seconds for the two units
to re-establish communication before running the STATIM.
9. The STATIM and Data Logger are now ready for use.
Data Logger Specifications
Connection: Serial Cable DB9F-DB25M
Baud Rate: 9600
Electrical Rating: 100V-240V, 0.6A, 50/60 Hz, multiplug
ACKNOWLEDGED MSD
USB Flash Drives:
Cruzer mini 1 GB
Memorex Travel drive 1 GB
Kingston Travel drive 512 MB
PNY Attache 1 GB
Lexar Jump Drive 1 GB
Other USB Flash Drives and SD Cards may also be compatible with this Data Logger.
SD Memory Cards:
Kingston 512 MB
Lexar 512 MB
PNY 512 MB
Ultra 512 MB
Kingston ultimate 512 MB

Table of Contents