Page 148
STAT
IM
2000/2000S Service Guide
96-108639 Rev 3.0
7. Electrical and Electronic Components
The chamber and steam generator
thermocouples MUST be recalibrated.
Do not calibrate a thermocouple until it
is properly installed and positioned in
the unit.
Reattach the pressure interface board,
(S class units only), or if a validation
thermocouple is present (later or
upgraded non S units) the validation
thermocouple interface board. Note
that on some non S units where no
validation thermocouple is present,
a printer cable is attached in this
socket. Replace this if present.
Rev.3.x/4.x units DO NOT have this
socket.
8. A dielectric strength test (Hi-Pot) and a protective bonding impedance test (ground continuity
should be performed on the STAT
IM
unit at this stage.
NOTE: These tests must be performed on the STATIM again once the work is completed and
the cover has been returned to the unit.
Controller Board Fuse Values
Always replace blown fuses with a fuse of the same size and value as indicated directly on the
controller board.