Page 123
STAT
IM
2000/2000S Service Guide
96-108639 Rev 3.0
6. Cover Assembly
STATIM 2000/2000S/2000 G4 Service Guide
Cover Assembly
For STATIM 2000 Classic
Removing the Cover
To remove the cover, proceed as follows (see Figure 1):
1. Inspect the cover (1) to ensure that it has not been damaged in transit. Position the unit on the
bench so the front leveler feet are securely on the work surface and the front cover extends
past the work surface edge.
2. Check that the LCD (2) and keypad (3) function. This requires that the unit be powered ON.
Check for symptoms of the failure to ensure that no further damage occurs.
3. Turn the power OFF and unplug the power cord (4) from the wall outlet and remove the
cassette from the unit.
4. Remove the reservoir cap (5) (and reservoir filter where fitted) from the top of the unit.
5. Using a No. 2 Phillips screwdriver, remove seven screws from the cover; four screws from the
sides and three at the rear of the unit.
6. If a bacteria-retentive air filter is fitted, unclip the filter from the filter bracket before attempting
to remove the cover.
Note: Before attempting the next step, be aware that there are cables connecting the
cover and the controller. If due care is not taken when removing the cover, damage may
occur.
7. With the fascia hanging over the edge of the work surface, lift the rear of the cover until it
clears the internal components, and then slide the cover forward about 3 cm / 1 inch to clear
the front of the armature and the unsupported controller board (8).
8. Gently lift the cover up slightly and pivot it around the front left corner of the unit, as shown
in Figure 1. When the cover is clear of the unit, rest it on its side. Take care not to strain the
ribbon cables connecting the display and keypad.
9. Observe the orientation of the ribbon cable connections. Disconnect the LCD and Keypad
connectors from the PCB.
Tip 1: To disconnect the keypad connector from the controller board, gently release the
lock (tab) on the connector housing of the controller board and pull down on the connector.
Tip 2: On some units there may be a cable tie securing the LCD cable connector to the
PCB. Check for this before attempting to pull off the connector. Remove the cable tie as
needed. A new cable tie is not required if the unit is not being shipped by courier.
10. Replace the reservoir cap to prevent unwanted materials and debris from entering the
reservoir.