Page 13
STAT
IM
2000/2000S Service Guide
96-108639 Rev 3.0
2. Tools, Maintenance Schedules, Procedures and Testing
STATIM 2000/2000S/2000 G4 Service Guide
Tools, Maintenance Schedules,
Procedures and Testing
Tools
STAT
IM
specific items from SciCan
Before a STAT
IM
can be serviced, the following special tools are required in addition to your
service tool kit. These tools are available from SciCan or your nearest service depot:
1. Control box SciCan Part # 01-103141S
2. Pump tester SciCan Part # 01-100713S
3. Water conductivity tester SciCan Part # 01-103139S
4. Solenoid plunger tube wrench SciCan Part # 01-103471S
5. 9/64” ball-end allen-key (hex) screwdriver SciCan Part # 01103469S
6. Calibration cassette,
SciCan 2000 Non S models only SciCan Part # 01-103087S
7. Calibration cassette,
SciCan 2000 All models SciCan Part # 01-106366S
Generic reference devices for calibration
8. Calibrated digital thermometer with ‘K’ type probe
9. Calibrated digital voltmeter with accuracy of 0.001 Volts
10. Calibrated digital pressure meter, 0 – 400 kPa absolute/0 – 7 bar absolute
More details on recommended devices can be found in section ‘Recommended Reference Meters’
Electrical safety test devices
11. Hi-Pot tester
12. Ground continuity tester
13. Static strap
14. Static bags
General tool list
t Phillips screwdriver
t Flat-blade electrician’s screwdriver
t Potentiometer trimmer
t Needle nose pliers
t Wire cutters
t Wrench 3/8” A/F
t Wrench 7/16” A/F
t Wrench 9/16” A/F
t Wrench 11/16” A/F
t Wrench 7/8” A/F
Note: This tool list is a guide and suggests what is the minimum required to remove and replace
components in the STATIM. Other tools may be required as an aid to servicing at the discretion of
the individual service engineer.