Page 131
STAT
IM
2000/2000S Service Guide
96-108639 Rev 3.0
6. Cover Assembly
STATIM 2000/2000S/2000 G4 Service Guide
For STATIM 2000 G4
Removing the STAT
IM
2000 G4 Cover
To remove the cover, follow these steps (see Figure 5):
1. Inspect the cover (1) to ensure that it has not been damaged in transit. Position the unit on the
bench so the front leveller feet are securely on the work surface and the front cover extends
past the work surface edge.
2. Check that the colour LCD functions. This requires that the unit be powered ON
Note: Check for symptoms of the failure to ensure that no further damage occurs.
3. After checking the LCD, turn the unit OFF, unplug the power cord from the unit’s line filter (2)
and remove the cassette from the unit.
4. Remove the reservoir cap (3) from the top of the unit.
5. Remove seven screws from the cover; first remove two on each side (4) and then three with
lockwashers (5) at the rear of the unit, using a #2 Phillips screwdriver. The Air Filter (6) must be
disconnected from the bracket before the cover can be removed.
6. Slowly, lift the rear portion of the cover upwards. When the cover is clear of the internal
components, carefully slide the entire cover forward to clear the front of the armature and of
the unsupported controller board (8). Slowly pivot the cover the same way a book opens and
place it next to the unit. The colour LCD should still be accessible and the internal components
exposed.
Note: The controller board is unsupported at the front of the unit. Be careful. The LCD cable
and LCD DC power source cable are still attached to the unit.
7. Disconnect the LCD connector (8) from controller board header P3, and the LCD DC power
source cable (9) from the DC power source (10) located at the back of the unit.
8. Replace the reservoir cap onto the top of the reservoir.