Page 34
STAT
IM
2000/2000S Service Guide
96-108639 Rev 3.0
2. Tools, Maintenance Schedules, Procedures and Testing
Routine Maintenance Schedule
and Procedures
To maintain the STATIM in good order between annual services, follow these guidelines:
DAILY
Water Reservoir
t Replace water as needed.
t For opthalmic use, drain at the end of every workday, leave empty, and refill at the start of
the next workday.
Water Bottle
t Empty the waste bottle every time you refill the reservoir.
t Fill with water, up to MIN line marking. You may also add some chlorine-free disinfectant.
WEEKLY
Cassette
t Wash the interior of the cassette with dishwashing soap or a mild detergent that does not
contain chlorine.
t Scrub the inside with a cleaning pad designed for use with Teflon™-coated surfaces.
t After removing all traces of the detergent, treat interior surfaces of the cassette with the
STAT-DRI™ Plus drying agent to enhance the drying process. Order more STAT-DRI™ Plus
from SciCan quoting 2OZPLUS, 8OZPLUST, or 32OZPLUS. Please note that STAT-DRI
drying agent is not be to used with the U.S. G4 models (G4-121101 and G4-201103).
Biological and/or Air Filter
t Check the filter for dirt and moisture. Replace if dirty. Call for service if wet.
Water Reservoir Filter
t Check the water reservoir filter every week and clean if necessary. Replace only if
necessary.
EVERY 6 MONTHS
Cassette Seal
t Replace every 500 cycles or six months (whichever is first), or whenever necessary.
Biological and/or Air Filter
t Replace every 500 cycles or six months (whichever is first).
STATIM 2000/2000S/2000 G4 Service Guide

Table of Contents