Page 45
STAT
IM
2000/2000S Service Guide
96-108639 Rev 3.0
2. Tools, Maintenance Schedules, Procedures and Testing
STATIM 2000/2000S/2000 G4 Service Guide
Water Conductivity Measurement
Figure 26
Water Conductivity Testing
Water conductivity testing determines the amount of
dissolved solids in the steam-process distilled water used
in the STAT
IM
unit. Use only steam-process distilled water
having 5 ppm or less dissolved solids or a conductivity of
less than 10 S / cm. Follow the manufacturer’s instructions
to test water conductivity using SciCan conductivity meter,
part #01-103139S, or any other equivalent Meter.
Temperature has a significant effect on conductivity
readings, therefore the water being tested should be at
room temperature.