8. Water Pumps, Reservoir, and Compressor
STATIM 2000/2000S/2000 G4 Service Guide
Reinstalling the Fluid-O-Tech Pump Assembly
To reinstall the pump assembly, proceed as follows (see Figure 14):
1. Slide the outlet side of the pump under the solenoid valve bracket, locate the pump assembly
into the mounting bracket and snap it into place. Do not kink the Teflon™ tube (9).
2. Push the open end of the rubber inlet tube (6) extending from the water reservoir on to the
pump inlet fitting (7) as far as it will go. Secure the tube to the fitting using a cable tie (4).
3. Attach the compression nut (8) on the Teflon™ tube to the inlet fitting (11) on top of the steam
generator (10) finger tight. Tighten this nut using a 3/8-inch wrench. Do not overtighten.
4. Re-attach the diode plug (2) using the diode plug screw (1).
5. A dielectric strength test (Hi-Pot) and a protective bonding impedance test (ground continuity)
should be performed on the STAT
IM
unit at this stage.
NOTE: These tests must be performed on the STATIM again once the work is completed and
the cover has been returned to the unit.
6. Make sure there is sufficient steam-process distilled water in the reservoir and activate the
pump using the control box for 5 seconds to ensure that the pump is functioning.
7. Fasten one cable tie to each end of the pump mounting bracket.
8. Run a sterilization cycle and observe all fittings and tubes for leaks. Check LCD read-outs for
messages indicating cycle status.