Page 250
STAT
IM
2000/2000S Service Guide
96-108639 Rev 3.0
12. Printer and Data Logger
12. Using the UNWRAPPED and WRAPPED buttons, scroll until the cursor is next to “Save and
Exit” and press the RUBBER & PLASTICS button, which will now save the data, and take you
out of the user menu and back to the operating menu.
Note: When printed or electronic records are being kept for future reference, it is important to
ensure the correct data is printed, and a check should be undertaken at this stage to ensure that
the date, time and unit number (where appropriate) are set before using a data recording device.
This may have been undertaken during installation, if not, refer to the user manual or to this service
manual’s Chapter 13, “Using the service menu” for information on resetting the time, date and unit
number.
13. Power OFF the STATIM unit.
Connecting the Printer
1. Connect the printer to the STATIM unit using a Serial Null Modem Cable DB9F – DB25M and
secure the cable.
2. Connect the printer power supply to the mains power socket as appropriate to the printer being
used.
3. Power ON the printer.
4. Power ON the STATIM unit.
5. The unit is now ready for use.
Data Logger
SciCan’s Data Logger can record and store cycle information onto a mass storage device (MSD)
such as a USB Flash Drive or SD memory card.
The Data Logger is supplied with two cables for connection to
tQJODPNNVOJDBUJPOQPSUTSTATIM 2000 units with revision 2.x/5.x/6.x controller
boards only) and
tQJODPNNVOJDBUJPOQPSUTBMMBOEVOJUTXJUISFWJTJPOYDPOUSPMMFS
boards)
Notes:
tSTATIM 2000 units can only be fitted with a data logger if a communication port is
present. Early STATIM 2000 units (with revision 3.x/4.x controller boards) that are
NOT fitted with a port can not have a port retrofitted.
tSTATIM 2000 units that have been upgraded to a revision 7 controller board will
require the 25 pin port upgrading to a 9 pin port to ensure correct communication.
A kit is available to enable this which is RS232 Port Kit STATIM 2000/2000S part
number 01-110221S.
STATIM 2000/2000S/2000 G4 Service Guide