Page 48
STAT
IM
2000/2000S Service Guide
96-108639 Rev 3.0
3. Troubleshooting Cycle Faults
3. Turn the STATIM ON. The unit is now in calibrating mode.
4. For version 2 and 3 software without selecting a cycle, press the START button. For other
versions of software press the “Wrapped” button.
5. The printer will print the last complete or incomplete cycle. This information can be useful in
diagnosing intermittent problems when the cycle fault code has not been recorded.
For units with revision 7.x controllers
(with service menu), including G4
All revision units fitted with revision 7.x controllers have a service menu which is designed to:
t1SPWJEFJOGPSNBUJPOPOUIFTFUVQPGUIFNBDIJOF
t"MMPXUIFDIBOHJOHPSSFTFUUJOHPGWBSJPVTNBDIJOFQBSBNFUFST
t0QFSBUFWBSJPVTBTQFDUTPGUIFVOJUXJUIUIFDPWFSPOBTPQQPTFEUPVTJOHBDPOUSPMCPY
Certain information provided in this menu may be used to assist in diagnostics.
Using the service menu on the STATIM Classic
(Rev. 7 controller board)
The service menu can be displayed on the LCD screen of any STATIM Classic with a Rev. 7 board
by leaving the cover on, or using the LCD screen of the control box if the cover is removed and
detached from the unit.
t5PFOUFSUIF4FSWJDF.FOVQPXFS0/UIF45"5IM while holding down the “Unwrapped”
and “Wrapped” keys.
t:PVXJMMUIFOCFQSPNQUFEUPFOUFSUIFTFSWJDFQBTTXPSEUPBMMPXZPVUPFOUFSUIF
service menu.
t5IFEFGBVMUQBTTXPSEJTFOUFSFECZQSFTTJOHUIFi6OXSBQQFEwi8SBQQFEwi3VCCFSBOE
Plastics” and “Stop” keys in that order.
t5IFTFSWJDFUFDIOJDJBODBODIBOHFUIJTQBTTXPSE*ODBTFUIFDIBOHFEQBTTXPSEJT
lost or has been changed by another technician, a backdoor password can be used
(which can not be changed) by pressing the “Unwrapped”, “Wrapped”, “Unwrapped”,
“Wrapped” keys in this order.
STATIM 2000/2000S/2000 G4 Service Guide