EasyManua.ls Logo

SciCan Statim 2000 - For Revision 3.X and 4.X Controllers

SciCan Statim 2000
275 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 66
STAT
IM
2000/2000S Service Guide
96-108639 Rev 3.0
3. Troubleshooting Cycle Faults
STATIM 2000/2000S/2000 G4 Service Guide
Check
Cassette
#6
The steam generator temperature is more
than 5°C higher than the chamber
temperature.
Most probable causes are a blocked duct, a blocked solenoid valve, miscalibration or a pinched
exhaust tube.
Check
Cassette
#7
The cassette temperature has dropped
below 130.5°C during the sterilizing phase
of the UNWRAPPED or WRAPPED Cycle,
or below 117.5°C during the sterilizing
phase of the RUBBER & PLASTIC
Cycle.
Most probable cause is a faulty solenoid valve.
Check
Cassette
#8
The chamber temperature is more than 5°C
higher than the steam generator
temperature.
Most probable causes are thermocouple miscalibration or very strong pump delivery.
Check
Cassette
#9
This is a special error message for the U.K.
market. The chamber temperature has
exceeded 139°C during the UNWRAPPED/
WRAPPED Cycle, or has exceeded 126°C
during the RUBBER & PLASTIC Cycle.
Most probable causes are a blocked or faulty solenoid valve or a pinched exhaust tube.
Service
Needed
The steam generator temperature has
exceeded 170°C, chamber temperature has
exceeded 147°C, or the thermocouple is
faulty.
Most probable causes are a blocked or faulty solenoid valve, a pinched exhaust tube, a broken
thermocouple lead, an open, disconnected or defective thermocouple, a weak pump or a
contaminated steam generator.
Check
Cassette
(without a
number)
The cassette temperature has failed to drop
to 103°C within a timeout period at the end
of a cycle.
Most probable causes are a blocked exhaust duct in the cassette, a blocked or faulty solenoid
valve or a pinched / kinked exhaust tube.
CF
number
Description of Error Message
Probable Cause of Error Message
Check
Cassette
#1
The cassette temperature failed to reach
95°C in 3 minutes.
This error message occurs only during warm up. The most probable causes are:
1. Steam generator does not heat up. No power to steam generator. Blown thermal fuse.
2. Check for a large cassette leak or a large load.
Check
Cassette
#2
The cassette temperature failed to increase
from 95°C to 100°C within 1 minute 20
seconds.
Most probable causes are a faulty cassette seal, miscalibration, a damaged cassette, improperly
adjusted or failed microswitch allowing unit to function when cassette is not fully inserted.
Occasionally, a low pressure weather system in higher altitude sites will alter the boiling
temperature significantly and trigger this fault.
Verify Vref to be 4.0V.
Check
Cassette
#3
The cassette has failed to pressurize and
achieve a temperature of 110°C within 70
seconds of pressurization.
This may be caused by a faulty or worn cassette seal, a faulty or dirty solenoid valve, a damaged
cassette or a faulty exhaust tube, leaky check valve or pressure relief valve.
Check
Cassette
#4
The cassette has failed to achieve
sterilization conditions within 10 minutes of
the chamber first reaching
102°C.
Several conditions may cause this: a damaged or worn cassette seal, a faulty solenoid valve,
improperly installed copper tubing, cassette thermocouple misalignment or damage, a faulty
steam generator pressure relief valve or a faulty check valve on.
Check
Cassette
#5
The software causes the pump to activate
while between 2 minutes 44 seconds and 2
minutes 24 seconds remaining in the cycle.
If a request to pump water occurs outside of
2 minutes 44 seconds, Check Cassette #5
occurs.
This indicates a substantial steam leak from the cassette, a contaminated steam generator or a
weak pump.
Cycle Fault Numbers and descriptions on units with revision 3.x and 4.x type controller boards

Table of Contents

Other manuals for SciCan Statim 2000

Related product manuals