2000/2000S Service Guide
96-108639 Rev 3.0
5. STAT
IM
2000S Calibration
t IMPORTANT – now re-check the actual temperature reading (CCC.C) with the
reading on the temperature reference meter. The readings need to be stable, so allow
sufficient time to compare them.
t If you are confident that the temperature readings are still within ±0.25°C, then the unit
is in equilibrium and the calibration has finished.
t If the readings are outside of the ± 0.25°C tolerance then further diagnostics may be
required prior to re-calibration.
10. Press the STOP button on the control box when finished.
3. Switch off the unit and remove all calibration equipment.
STATIM 2000/2000S/2000 G4 Service Guide