2000/2000S Service Guide
96-108639 Rev 3.0
7. Electrical and Electronic Components
6. A dielectric strength test (Hi-Pot) and a protective bonding impedance test (ground continuity)
should be performed on the STAT
IM
unit at this stage.
NOTE: These tests must be performed on the STATIM again once the work is completed and
the cover has been returned to the unit.
7. Plug the power cord (3) into the line filter and the wall receptacle and turn the power switch (2)
ON. Observe the LCD and indicator lights to determine that power is present.