3. Troubleshooting Cycle Faults
STATIM 2000/2000S/2000 G4 Service Guide
Important Note: Any changes made to the chamber and voltage calibration offsets and
conductivity settings will be reset.
The cycle counter will not be reset.
The default setting on all STATIM units is NO.
The alternative is YES, RESET NVRAM.
To activate a reset, proceed as follows as follows:
1. The UNWRAPPED or WRAPPED keys will switch the screen display to “YES, RESET NVRAM”.
2. The RUBBER & PLASTICS key will action your selection and return the unit to the service menu
condition.
3. Pressing the STOP key will exit without changes and return the unit to the operational menu.
>Change Password
This function will allow you to change the default password (Unwrapped, Wrapped, Rubber and
Plastics, Stop) to a new key/button sequence.
When entering this sub menu, the unit will prompt you as follows:
t5IFVOJUXJMMBTLZPVUPFOUFSBOFXLFZQBTTXPSE
t&OUFSUIFOFXQBTTXPSE
t5IFVOJUXJMMUIFOBTLZPVUPSFFOUFSUIFOFXQBTTXPSE
t3FFOUFSUIFQBTTXPSE
t5IFVOJUXJMMDPOmSNUIBUUIFQBTTXPSEIBTCFFODIBOHFEPSJGDIBOHJOHUIFQBTTXPSE
failed, the unit will again ask you to enter a new 4-key password.
>Backup NVRAM
Screen display when in Backup NVRAM sub menu.
This function will allow the backup of any revised unit settings after calibration for example.
The default setting on all STATIM units is NO.
The alternative is YES.
Type New Password
****
Retype New Password
****
Password Changed
>Backup NVRAM
No